title: What should I check first if I get an unexpected CM reading?
description: A comprehensive troubleshooting procedure for an unexpected reading on your CM (Ionic Contamination Test Results)
---

[Skip to content](https://help.gen3systems.com/techshot/cm-check-procedure#main-content)

English

Show submenu for translations

[Need more support?](https://help.gen3systems.com/techshot/kb-tickets/new?hsLang=en)

[![Orange white GEN3 logo-1](https://help.gen3systems.com/hs-fs/hubfs/Orange%20white%20GEN3%20logo-1.png?width=250&height=104&name=Orange%20white%20GEN3%20logo-1.png)](https://www.gen3systems.com/)

Open main navigation

Close main navigation

- English
  
  Show submenu for translations
- [Need more support?](https://help.gen3systems.com/techshot/kb-tickets/new)
- [Contact us](https://www.gen3systems.com/tech-shot)

[Contact us](https://www.gen3systems.com/tech-shot)

 How can we help you?

- There are no suggestions because the search field is empty.

1. [TechShot](https://help.gen3systems.com/techshot?hsLang=en)
2. [Troubleshooting](https://help.gen3systems.com/techshot/troubleshooting?hsLang=en)
3. [CM Series - Ionic Contamination Testing](https://help.gen3systems.com/techshot/troubleshooting?hsLang=en#cm-series-ionic-contamination-testing)

August 13, 2026

# What should I check first if I get an unexpected CM reading?

## A practical troubleshooting sequence for your CM

**Step 1 — Temperature**  
Confirm that the test is being performed within the recommended 24°C–30°C range.

**Step 2 — Solution**  
Confirm the correct water/IPA ratio and that the materials meet the recommended grade.

**Step 3 — Solution condition**  
Consider its age, exposure to air and possible IPA evaporation.

**Step 4 — Sample-to-solution ratio**  
Confirm the recommended 10 ml/cm² ratio.

**Step 5 — Test duration**  
Confirm that the test has not exceeded 15 minutes.

**Step 6 — Environment**  
Consider whether airborne contamination, temperature variation or an uncontrolled factory environment could be influencing the result.

**Step 7 — Calibration**  
Use the correct calibration solution and the established multi-test averaging procedure.

- [FAQs](https://help.gen3systems.com/techshot/faqs?hsLang=en#main-content)

    - [CM Series - Ionic Contamination Testing](https://help.gen3systems.com/techshot/faqs?hsLang=en#cm-series-ionic-contamination-testing)
    - [AutoSIR2+ Surface Insulation Resistance Testing](https://help.gen3systems.com/techshot/faqs?hsLang=en#autosir2-surface-insulation-resistance-testing)
- [Getting started](https://help.gen3systems.com/techshot/getting-started?hsLang=en#main-content)

    - [CM Series - Ionic Contamination Testing](https://help.gen3systems.com/techshot/getting-started?hsLang=en#cm-series-ionic-contamination-testing)
    - [MUST3](https://help.gen3systems.com/techshot/getting-started?hsLang=en#must3)
- [Troubleshooting](https://help.gen3systems.com/techshot/troubleshooting?hsLang=en#main-content)

    - [MUST3 - Solderability Testing](https://help.gen3systems.com/techshot/troubleshooting?hsLang=en#must3-solderability-testing)
    - [CM Series - Ionic Contamination Testing](https://help.gen3systems.com/techshot/troubleshooting?hsLang=en#cm-series-ionic-contamination-testing)
- [General](https://help.gen3systems.com/techshot/general?hsLang=en)
- [Objective Evidence](https://help.gen3systems.com/techshot/objective-evidence?hsLang=en#main-content)

    - [Objective Evidence](https://help.gen3systems.com/techshot/objective-evidence?hsLang=en#objective-evidence)

[![Shield-Orange](https://help.gen3systems.com/hs-fs/hubfs/Branding/Shield-Orange.png?width=100&height=100&name=Shield-Orange.png "Shield-Orange")](https://www.gen3systems.com/)

B2, ARMSTRONG MALL, SOUTHWOOD BUSINESS PARK, FARNBOROUGH, HAMPSHIRE, GU14 0NR

<https://www.facebook.com/> <https://www.twitter.com/> <https://www.instagram.com/> <https://podcasts.apple.com/> [mailto:email@email.com](mailto:email@email.com)

Copyright © 2026, GEN3